Current

SCPI Commands

FETCh:WCDMa:MEASurement<Instance>:MEValuation:TRACe:PHD:CURRent
READ:WCDMa:MEASurement<Instance>:MEValuation:TRACe:PHD:CURRent
class Current[source]

Current commands group definition. 2 total commands, 0 Sub-groups, 2 group commands

fetch()List[float][source]
# SCPI: FETCh:WCDMa:MEASurement<instance>:MEValuation:TRACe:PHD:CURRent
value: List[float] = driver.multiEval.trace.phd.current.fetch()

Returns the values of the phase discontinuity traces for up to 120 slots. One value per measured slot is returned (see method RsCmwWcdmaMeas.Configure.MultiEval.msCount) .

INTRO_CMD_HELP: The meaning of the value depends on the measurement period (see method RsCmwWcdmaMeas.Configure.MultiEval.Mperiod.modulation) :

  • For full-slot measurements, each value indicates the phase discontinuity at the boundary between a slot and the previous slot. As there is no previous slot for slot 0, the first returned phase discontinuity value equals NCAP.

  • For half-slot measurements, each value indicates the phase discontinuity at the boundary between the first and second half-slot of a slot. This value can be measured for all slots, including slot 0.

See also ‘Detailed Views: Phase Discontinuity’

Use RsCmwWcdmaMeas.reliability.last_value to read the updated reliability indicator.

return

phase_disc: float One value per measured slot Range: -180 deg to 180 deg, Unit: deg

read()List[float][source]
# SCPI: READ:WCDMa:MEASurement<instance>:MEValuation:TRACe:PHD:CURRent
value: List[float] = driver.multiEval.trace.phd.current.read()

Returns the values of the phase discontinuity traces for up to 120 slots. One value per measured slot is returned (see method RsCmwWcdmaMeas.Configure.MultiEval.msCount) .

INTRO_CMD_HELP: The meaning of the value depends on the measurement period (see method RsCmwWcdmaMeas.Configure.MultiEval.Mperiod.modulation) :

  • For full-slot measurements, each value indicates the phase discontinuity at the boundary between a slot and the previous slot. As there is no previous slot for slot 0, the first returned phase discontinuity value equals NCAP.

  • For half-slot measurements, each value indicates the phase discontinuity at the boundary between the first and second half-slot of a slot. This value can be measured for all slots, including slot 0.

See also ‘Detailed Views: Phase Discontinuity’

Use RsCmwWcdmaMeas.reliability.last_value to read the updated reliability indicator.

return

phase_disc: float One value per measured slot Range: -180 deg to 180 deg, Unit: deg