Current¶
SCPI Commands
READ:WCDMa:MEASurement<Instance>:MEValuation:PCDE:CURRent
FETCh:WCDMa:MEASurement<Instance>:MEValuation:PCDE:CURRent
- class Current[source]
Current commands group definition. 2 total commands, 0 Sub-groups, 2 group commands
- class ResultData[source]
Response structure. Fields:
Reliability: int: decimal ‘Reliability Indicator’
Pcd_Error: float: float Peak code domain error Range: -100 dB to 0 dB, Unit: dB
Pcd_Error_Phase: enums.PcdErrorPhase: IPHase | QPHase Phase where the peak code domain error was measured IPHase: I-Signal QPHase: Q-Signal
Pcd_Error_Code_Nr: int: decimal Code number for which the PCDE was measured Range: 0 to 255
- fetch() → ResultData[source]
# SCPI: FETCh:WCDMa:MEASurement<instance>:MEValuation:PCDE:CURRent value: ResultData = driver.multiEval.pcde.current.fetch()
Returns the peak code domain error (PCDE) results. In addition to the current PCDE value, the maximum PCDE value can be retrieved. See also ‘Detailed Views: CD Monitor’
- return
structure: for return value, see the help for ResultData structure arguments.
- read() → ResultData[source]
# SCPI: READ:WCDMa:MEASurement<instance>:MEValuation:PCDE:CURRent value: ResultData = driver.multiEval.pcde.current.read()
Returns the peak code domain error (PCDE) results. In addition to the current PCDE value, the maximum PCDE value can be retrieved. See also ‘Detailed Views: CD Monitor’
- return
structure: for return value, see the help for ResultData structure arguments.