Maximum

SCPI Commands

READ:WCDMa:MEASurement<Instance>:MEValuation:CARRier<Carrier>:TRACe:UEPower:MAXimum
FETCh:WCDMa:MEASurement<Instance>:MEValuation:CARRier<Carrier>:TRACe:UEPower:MAXimum
class Maximum[source]

Maximum commands group definition. 2 total commands, 0 Sub-groups, 2 group commands

fetch(carrier=<Carrier.Default: -1>)List[float][source]
# SCPI: FETCh:WCDMa:MEASurement<instance>:MEValuation:CARRier<carrier>:TRACe:UEPower:MAXimum
value: List[float] = driver.multiEval.carrier.trace.uePower.maximum.fetch(carrier = repcap.Carrier.Default)

Returns the values of the UE power traces for up to 120 slots. Each current value is averaged over a half-slot or a full-slot, depending on the measurement period (see method RsCmwWcdmaMeas.Configure.MultiEval.Mperiod.modulation) . The number of results depends on the measurement length (see method RsCmwWcdmaMeas.Configure.MultiEval.msCount) . The results of the current, average, minimum, maximum and standard deviation traces can be retrieved. The minimum and standard deviation trace cannot be displayed at the GUI. See also ‘Detailed Views: UE Power and Power Steps’

Use RsCmwWcdmaMeas.reliability.last_value to read the updated reliability indicator.

param carrier

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Carrier’)

return

ue_power: No help available

read(carrier=<Carrier.Default: -1>)List[float][source]
# SCPI: READ:WCDMa:MEASurement<instance>:MEValuation:CARRier<carrier>:TRACe:UEPower:MAXimum
value: List[float] = driver.multiEval.carrier.trace.uePower.maximum.read(carrier = repcap.Carrier.Default)

Returns the values of the UE power traces for up to 120 slots. Each current value is averaged over a half-slot or a full-slot, depending on the measurement period (see method RsCmwWcdmaMeas.Configure.MultiEval.Mperiod.modulation) . The number of results depends on the measurement length (see method RsCmwWcdmaMeas.Configure.MultiEval.msCount) . The results of the current, average, minimum, maximum and standard deviation traces can be retrieved. The minimum and standard deviation trace cannot be displayed at the GUI. See also ‘Detailed Views: UE Power and Power Steps’

Use RsCmwWcdmaMeas.reliability.last_value to read the updated reliability indicator.

param carrier

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Carrier’)

return

ue_power: No help available