Modulation

SCPI Commands

CONFigure:WCDMa:MEASurement<Instance>:MEValuation:LIST:SEGMent<Segment>:MODulation
class Modulation[source]

Modulation commands group definition. 1 total commands, 0 Sub-groups, 1 group commands

class ModulationStruct[source]

Structure for setting input parameters. Fields:

  • Mod_Statistics: int: integer The statistical length is limited by the length of the segment (see [CMDLINK: CONFigure:WCDMa:MEASi:MEValuation:LIST:SEGMentno:SETup CMDLINK]) . Range: 1 to 1000

  • Enable_Ue_Power: bool: OFF | ON OFF: Disable measurement ON: Enable measurement of UE power

  • Enable_Evm: bool: OFF | ON Disable or enable measurement of EVM

  • Enable_Mag_Error: bool: OFF | ON Disable or enable measurement of magnitude error

  • Enable_Phase_Err: bool: OFF | ON Disable or enable measurement of phase error

  • Enable_Freq_Error: bool: OFF | ON Disable or enable measurement of frequency error

  • Enable_Iq: bool: OFF | ON Disable or enable measurement of I/Q origin offset and imbalance

get(segment=<Segment.Default: -1>)ModulationStruct[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:LIST:SEGMent<nr>:MODulation
value: ModulationStruct = driver.configure.multiEval.listPy.segment.modulation.get(segment = repcap.Segment.Default)

Defines the statistical length for the AVERage, MAXimum, and SDEViation calculation and enables the calculation of the different modulation results in segment no. <no>; see ‘Multi-Evaluation List Mode’. The statistical length for CDP, CDE and modulation results is identical (see also method RsCmwWcdmaMeas.Configure.MultiEval.ListPy.Segment.CdPower.set) .

param segment

optional repeated capability selector. Default value: Nr1 (settable in the interface ‘Segment’)

return

structure: for return value, see the help for ModulationStruct structure arguments.

set(structure: RsCmwWcdmaMeas.Implementations.Configure_.MultiEval_.ListPy_.Segment_.Modulation.Modulation.ModulationStruct, segment=<Segment.Default: -1>)None[source]
# SCPI: CONFigure:WCDMa:MEASurement<instance>:MEValuation:LIST:SEGMent<nr>:MODulation
driver.configure.multiEval.listPy.segment.modulation.set(value = [PROPERTY_STRUCT_NAME](), segment = repcap.Segment.Default)

Defines the statistical length for the AVERage, MAXimum, and SDEViation calculation and enables the calculation of the different modulation results in segment no. <no>; see ‘Multi-Evaluation List Mode’. The statistical length for CDP, CDE and modulation results is identical (see also method RsCmwWcdmaMeas.Configure.MultiEval.ListPy.Segment.CdPower.set) .

param structure

for set value, see the help for ModulationStruct structure arguments.

param segment